AII GPR 1600/2600/3100 Oxygen Analyzers for Industrial Gas

Percent and PPM O2 Monitoring in Bulk Gases

Features:

  • High accuracy ( ±0.02 ppm in lowest range)
  • 24 months sensor life with normal usage
  • Sample in CO2 backgrounds with XLT sensor
  • Four measurement ranges with manual or auto-ranging
  • Two Alarms and range indication output
  • Bench, rack or wall mounting options
  • Integrated bypass valve (for GPR-1600)
  • Easy access and sensor replacement
  • Simple to use HMI

Our range of oxygen analyzers are designed to monitor gas produced or used in industrial processes where trace oxygen from low parts per million to pure O2 must be precisely measured. Utilizing AII’s high performance galvanic oxygen sensors these advanced instruments are simple to use with a common HMI and menu structure across-the-range chassis. Operators only need to learn just one instrument for multiple gas analysis applications at various oxygen levels. Available as bench-, 19" rack-, panel- and wall-mounted for flexible installation.

Specification:


GPR-1600
GPR-2600 GPR-3100
Measurement range 0-10, 0-100, 0-1000 ppm, 0-1%
(0-25% calibration only)
0-1%, 0-5%, 0-10%, 0-25% 0-100%, plus suppressed zero ranges.
50, 80, or 90-100%
Accuracy < ±2% of range
±0.1% at constant conditions after calibration with 95-100% oxygen (GPR-3100)
< ±2% of range
±0.1% at constant conditions after calibration with 95-100% oxygen (GPR-3100)
< ±2% of range
±0.1% at constant conditions after calibration with 95-100% oxygen (GPR-3100)
Response time T90 < 10 seconds T90 < 10 seconds T90 < 13 seconds
Recovery Time 60 sec in air to < 10 ppm in < 1 hour on N2 purge N/A N/A
Sensitivity (LDL) 50 ppb 50 ppm 0.1% oxygen
Linearity < 0.5% of scale < 0.5% of scale < 0.5% of scale
Sensor model GPR-12-333

XLT-12-333 for gas mixture with > 0.5% CO2
GPR-11-32-4

XLT-11-24-4 for gas mixture with > 0.5% CO2
GPR-11-120-OP
Sensor life at 25°C and 1 atm 24 months in < 1000 ppm O2 GPR-11-32-4 32 months; XLT-11-24-4 24 months 24 months in 100% oxygen
Calibration interval Typically: 1-3 months Typically: 1-3 months Typically: 1-3 months
Inlet pressure 0.34 - 2 barg (5-30 psig) with atmospheric vent 0.34 - 2 barg (5-30 psig) with atmospheric vent 248 bar g (3600psig)
Flow rate 0.5 - 1.0 l/m (1-2 SCFH) 0.5 - 1.0 l/m (1-2 SCFH) 0.5 - 1.0 l/m (1-2 SCFH)
Gas connections 1/4" compression tube fittings 1/4" compression tube fittings 1/4" compression tube fittings
Display Graphical LCD 12.7 x 7cm (5 x 2.75"); resolution 0.01 Graphical LCD 12.7 x 7cm (5 x 2.75"); resolution 0.01 Graphical LCD 12.7 x 7cm (5 x 2.75"); resolution 0.01
Enclosure Painted aluminum
See individual mounting options for dimensions
Painted aluminum
See individual mounting options for dimensions
Painted aluminum
See individual mounting options for dimensions
Compensation Barometric pressure and temperature;
Temperature controlled sample system and sensor (GPR-3100 only)
Barometric pressure and temperature;
Temperature controlled sample system and sensor (GPR-3100 only)
Barometric pressure and temperature;
Temperature controlled sample system and sensor (GPR-3100 only)
Analog output 4-20 mA isolated, 0-1V, and 0-5V 4-20 mA isolated, 0-1V, and 0-5V 4-20 mA isolated, 0-1V, and 0-5V
Range ID 1-5 V or 4-20mA, optional relay contacts 1-5 V or 4-20mA, optional relay contacts 1-5 V or 4-20mA, optional relay contacts
Communications Choose from RS485, RS232 or USB Choose from RS485, RS232 or USB Choose from RS485, RS232 or USB
Alarms Two user-adjustable alarm relays Two user-adjustable alarm relays Two user-adjustable alarm relays
Operating temperature GPR sensor: 5°C to 45°C (41°F to 113°F)
XLT sensor: -10°C to 45°C (14°F to 113°F)
GPR sensor: 5°C to 45°C (41°F to 113°F)
XLT sensor: -10°C to 45°C (14°F to 113°F)
GPR sensor: 5°C to 45°C (41°F to 113°F)
XLT sensor: -10°C to 45°C (14°F to 113°F)
Power Universal 100-240 V AC Universal 100-240 V AC Universal 100-240 V AC